062AE 147A PJS 2/27/20 00207PJS2 SENSOR FOR SO2 LEAK DETECTOR REACTS SLOWLY <br /> DURING CALIBRATION AND TESTING.NEEDED TO BE <br /> CALIBRATED TWICE IN ORDER TO PASS.NEEDS A NEW <br /> SENSOR BEFORE IT FAILS COMPLETELY.New cell <br /> (A-ULTX-SENS-70-0-0)installed on 02/27/2020.Calibrated and <br /> • tested OK. <br /> 062FIT146C PJS 5/21/20 00521JAK1 BLINKING RED LIGHT ON PANEL/FRONT OF METER. In this <br /> • case the irradically blinking red LED indicates the detection of minute <br /> reverse movement of process fluid during zero flow when the reverse <br /> flow measurement configuration is turned off.There is no problem. <br /> HTR-6203 EC 1/3/20 91230DER1 HEATER IS NOT WORKING----ordered new heater, installed new <br /> heater and thermostat.Tested ok. <br /> PMP062SBSOI CLC 4/1/20 00309FJS1 PUMP#1 LEAK DETECTOR ALARM WILL NOT CLEAR.--- <br /> replaced sensor <br /> TIT-146A PJS 9/24/20 91211 PJS I RETROFIT SO2 LINE TEMPERATURE TRANSMITTERS TO <br /> REMOTE-MOUNT FOR MUCH EASIER ACCESS DURING <br /> CALIBRATIONS. PROBE AND TRANSMITTER LOCATED <br /> INSIDE SODIUM BISULFITE BUILDING.After researching costs <br /> of Robroy and labor it was determined the cost outweighed the <br /> benefit.The project was stopped. <br /> TIT-146A PJS 6/23/20 00618PJS1 DURING YEARLY CALIBRATION CHECK IT WAS NOTICED <br /> THE CALIBRATION WAS OFF QUITE A BIT,ABOUT 0.50mA <br /> ON THE TOP END.COULD NOT COMMUNICATE WITH THE <br /> LAPTOP TO CALIBRATE IT. Replaced transmitter,configured and <br /> calibrated. <br /> TIT-146B PJS 9/24/20 91211PJS2 RETROFIT SO2 LINE TEMPERATURE TRANSMITTERS TO <br /> REMOTE-MOUNT FOR MUCH EASIER ACCESS DURING <br /> • CALIBRATIONS. PROBE AND TRANSMITTER LOCATED <br /> INSIDE SODIUM BISULFITE BUILDING.After researching costs <br /> of Robroy and labor it was determined the cost outweighed the <br /> benefit.The project was stopped. <br /> 064 SODIUM HYPOCHLORITE FACILITY <br /> ITEM BY C-DATE W/O# <br /> 064LIT143C PJS 2/25/20 91212PJS1 HIGH LEVEL ALARM WHEN FILLING NEW TANK WITH <br /> CHLORINE,AND IT OCCURED WITH ONLY ABOUT 5 <br /> INCHES OF ACTUAL CHLORINE IN THE TANK. Looked at <br /> replacing the transmitter with an FMR52-N743/0,FMR62-VCR4/0, <br /> or FMR2O-C6U7/0(covering it with an Olefin bag to protect the <br /> plastic from UV rays)in case wiping the sensor head would not solve <br /> the problem. Wiping the sensor head fixed the problem-there was <br /> too much condensation inside the nearly empty tank. <br /> PMP-6402 JAM 4/7/20 00402DER2 THE DISCHARGE HOSE CAMLOC FITTINGS ARE LEAKING. <br /> ---Repaired. <br /> 075 COGENERATION FACILITY(NEW) <br /> ITEM BY C-DATE W/O# <br /> 075 MJN 10/26/20 01005GCP3 MAKE NEW BATTERY CABLE AND CONNECT BATTERY <br /> DISCONNECT SWITCH.---completed <br /> 076 COGENERATION FACILITY(OLD) <br /> ITEM BY C-DATE W/O# <br /> 076 MJN 6/10/20 90903BDW1 WO FOR PARTS PUCHASE,PLANNING AND ELECTRICAL <br /> • INSTALLATION OF TEMPORARY PACKAGE BOILER AT OLD <br /> COGEN BUILDING.----installation completed byEC electric and <br /> P <br /> Northwest Controls <br /> Page 15 of 20 <br />